Linear Calibration with Inverse Prediction, LOD and LOQ
Linear calibration fits instrument signal against standard concentrations and inverts the line to estimate unknown concentrations with confidence intervals, together with detection and quantitation limits.
선형 검량은 표준물질 농도에 대한 기기 신호를 적합하고 그 직선을 역으로 풀어 미지 시료 농도를 신뢰구간과 함께 추정하며 검출한계와 정량한계를 함께 구하는 방법입니다.
I measured standards and several unknown samples on my instrument; how do I compute the unknown concentrations with uncertainty and report LOD and LOQ?
표준물질과 미지 시료 몇 개를 기기로 측정했는데, 미지 농도를 불확도와 함께 계산하고 LOD·LOQ 를 어떻게 보고하나요?
Fit signal on concentration by ordinary or weighted least squares, then convert each unknown signal to concentration as (y0 - a)/b with a standard error that grows with distance from the centroid of the calibration. Report LOD and LOQ as 3.3 and 10 times the residual SD divided by the slope, or use blank replicates or Hubaux-Vos prediction bands, and check linearity with a lack-of-fit test when standards are replicated.
When to use Linear Calibration with Inverse Prediction, LOD and LOQ
- Unknown sample concentrations are read from a calibration line of standards.
- Method validation requires LOD, LOQ and linearity evidence (ICH Q2).
- Residual spread grows with concentration and weighting (1/x or 1/x²) is needed.
When to use something else
- The sample matrix changes the slope; use standard addition analysis.
- The response is clearly curved; use a quadratic or nonlinear calibration model.
Assumptions and requirements
- Concentrations of standards are known without appreciable error.
- Residuals are normal with constant variance (or variance matched by the chosen weights).
- The relationship is linear over the calibration range.
What the analysis reports
- Slope and intercept with SE and confidence intervals, s_y/x and R²
- Lack-of-fit F test when standards are replicated at three or more levels
- Inverse-predicted concentrations with SE and confidence intervals
- LOD and LOQ (ICH residual-SD, blank SD, or Hubaux-Vos)
- Residual plot
How to report it
The calibration line (n = [value]) was y = [value] + [value]x (s_y/x = [value], R² = [value]; lack of fit F([value], [value]) = [value], p = [value]); LOD = [value] and LOQ = [value], and the sample concentration was [value] (95% CI [value] to [value]).
Related methods
How this method is validated
이 방법의 검증 근거
- Published value. A test reproduces published reference values.
출판값. 출판된 참조값 재현 시험 1건 통과.
- Independent computation. 1 reference test compares the results with statsmodels.
독립 계산. statsmodels 대조 참조 시험 1건 통과.
Tolerances are relative differences (absolute differences for values near zero). R comparisons and reference tests are re-run before each release of the analysis engine.
허용오차는 상대 차이입니다(0 근처 값은 절대 차이). R 대조와 참조 시험은 분석 엔진을 배포할 때마다 다시 돌립니다.
Run Linear Calibration with Inverse Prediction, LOD and LOQ on Quorant
- Data you need
- a table (CSV or Excel) with an outcome variable and predictor (feature) columns
- Price
- 250 credits for one analysis up to 10,000 rows (complexity level 2); larger or wider data costs more. Optional methodologist review: 1200 credits.
- Checking
- Reproduces published reference values
- Before you pay
- A free pre-analysis diagnosis checks your data and gives an exact quote; nothing is deducted until you approve it.