Calibration Transfer Between Instruments (DS and PDS)
Calibration transfer by direct standardisation (DS) or piecewise direct standardisation (PDS) maps spectra from a slave instrument onto the master instrument's response so that a master calibration model can be reused.
직접 표준화(DS)·구간 직접 표준화(PDS)에 의한 검량선 이전은 보조 기기의 스펙트럼을 주 기기의 응답으로 변환해 주 기기의 검량 모형을 그대로 다시 쓰게 하는 방법입니다.
My NIR calibration was built on one spectrometer; how can I use it on a second instrument without rebuilding the model?
근적외선 검량 모형을 한 분광기로 만들었는데, 모형을 다시 만들지 않고 두 번째 기기에서 쓰려면 어떻게 해야 하나요?
Measure a small set of transfer samples on both instruments and estimate a transformation from slave to master spectra, either one full matrix (DS) or a banded matrix of local regressions within a wavelength window (PDS). Apply it to new slave spectra and check that spectral differences and prediction errors of the master model shrink on independent samples.
When to use Calibration Transfer Between Instruments (DS and PDS)
- The same samples can be measured on a master and a slave instrument.
- A master calibration model exists and rebuilding it on the slave would be costly.
- Instrument differences are mostly local shifts, broadening or intensity changes across channels.
When to use something else
- No paired transfer samples are available; use slope/bias correction on reference values or a robust global model.
- The instruments measure different spectral ranges or resolutions; resample first or rebuild the model.
Assumptions and requirements
- Spectral channels (up to 3,000) are the feature columns, the instrument is the group column, an optional sample id is the unit and optional reference values are the outcome.
- Transfer samples are measured on both instruments and paired by sample id or by order.
- Transfer samples span the spectral variation of future samples (selected here by Kennard-Stone).
- The instrument difference is stable over time.
What the analysis reports
- Spectral RMSE between slave and master before and after correction on test samples
- Transfer matrix summary and offset
- With a reference value column: RMSEP of a master PLS model on master, raw slave, corrected slave and slope/bias corrected predictions
- List of selected transfer samples
How to report it
PDS with [value] Kennard-Stone transfer samples and a window of [value] channels changed the RMSE between slave and master spectra from [value] to [value] on [value] test samples; RMSEP went from [value] to [value] (master [value]).
Related methods
- Kennard-Stone Calibration Set Selection
- Partial Least Squares Regression
- Spectral Preprocessing (SNV, MSC, Derivatives)
How this method is validated
이 방법의 검증 근거
- Independent computation. 1 reference test compares the results with direct computations (formulas, enumeration or simulation).
독립 계산. 직접 계산(식·열거·모의실험) 대조 참조 시험 1건 통과.
Tolerances are relative differences (absolute differences for values near zero). R comparisons and reference tests are re-run before each release of the analysis engine.
허용오차는 상대 차이입니다(0 근처 값은 절대 차이). R 대조와 참조 시험은 분석 엔진을 배포할 때마다 다시 돌립니다.
Run Calibration Transfer Between Instruments (DS and PDS) on Quorant
- Data you need
- a table (CSV or Excel) with predictor (feature) columns and a grouping variable
- Price
- 600 credits for one analysis up to 10,000 rows (complexity level 4); larger or wider data costs more. Optional methodologist review: 2400 credits.
- Checking
- Compared with independent computations
- Before you pay
- A free pre-analysis diagnosis checks your data and gives an exact quote; nothing is deducted until you approve it.