Grubbs Estimators of Instrument Imprecision
The method estimates how imprecise each of two or three instruments is when they all measure the same set of items and no reference standard exists. Because every instrument sees the same true values, the covariance between instruments estimates the variance of the items, and what remains of each instrument's own variance is its error variance. Standard errors follow from normal theory, the Pitman-Morgan test (two instruments) or pairwise Wald tests (three instruments) compare precisions, and paired t intervals estimate constant relative bias.
기준 표준이 없을 때 같은 품목들을 두세 개의 측정기기가 함께 잴 경우 각 기기가 얼마나 부정확한지를 추정합니다. 모든 기기가 같은 참값을 보므로 기기 사이의 공분산은 품목 자체의 분산을 추정하고, 각 기기의 분산에서 그것을 뺀 나머지가 그 기기의 오차분산입니다. 정규이론으로 표준오차를 구하고, 두 기기이면 Pitman-Morgan 검정, 세 기기이면 쌍별 Wald 검정으로 정밀도를 비교하며, 대응 t 구간으로 일정한 상대 편향을 추정합니다.
Two or three instruments measured the same specimens; which one is the more precise, and how large is each one's measurement error?
두세 개의 기기가 같은 시료를 측정했을 때, 어느 기기가 더 정밀하며 각 기기의 측정오차는 얼마나 큽니까?
The method returns each instrument's error variance with standard error and confidence interval, the item variance, reliability ratios, the test of equal precision and the mean differences between instruments with confidence intervals.
When to use Grubbs Estimators of Instrument Imprecision
- Several instruments, laboratories or raters measure the same items once each.
- No error-free reference measurement is available.
- You want separate error variances rather than only limits of agreement.
When to use something else
- You want limits of agreement between two methods: use Bland-Altman Agreement Analysis
- You need a calibration line between two error-prone methods: use Deming Regression
- You only want to compare the total variances of two paired measures: use Pitman-Morgan Test for Paired Variances
Assumptions and requirements
- Errors are independent of the true values and of each other
- Error variance is constant across items
- Approximate normality for the standard errors
What the analysis reports
- Error variance, SE and CI per instrument
- Item variance and reliability
- Test of equal precision
- Relative bias with CI
How to report it
Grubbs estimates from [value] items gave error variances of [value] (95% CI [[value], [value]]) for instrument A and [value] (95% CI [[value], [value]]) for instrument B; the Pitman-Morgan test of equal precision gave t([value]) = [value], p = [value].
Related methods
How this method is validated
이 방법의 검증 근거
- Independent computation. 2 reference tests compare the results with SciPy and direct computations (formulas, enumeration or simulation).
독립 계산. SciPy·직접 계산(식·열거·모의실험) 대조 참조 시험 2건 통과.
Tolerances are relative differences (absolute differences for values near zero). R comparisons and reference tests are re-run before each release of the analysis engine.
허용오차는 상대 차이입니다(0 근처 값은 절대 차이). R 대조와 참조 시험은 분석 엔진을 배포할 때마다 다시 돌립니다.
Run Grubbs Estimators of Instrument Imprecision on Quorant
- Data you need
- a table (CSV or Excel) with item columns (e.g. questionnaire items)
- Price
- 250 credits for one analysis up to 10,000 rows (complexity level 2); larger or wider data costs more. Optional methodologist review: 1200 credits.
- Checking
- Compared with independent computations
- Before you pay
- A free pre-analysis diagnosis checks your data and gives an exact quote; nothing is deducted until you approve it.