Single-Laboratory Precision Verification (CLSI EP15)
Single-laboratory precision verification (CLSI EP15) estimates repeatability and within-laboratory imprecision from replicates over several runs and checks them against the manufacturer's claims using upper verification limits.
단일 실험실 정밀도 검증(CLSI EP15)은 여러 날에 걸친 반복 측정으로 반복성과 실험실 내 정밀도를 추정하고, 검증 상한을 써서 제조사 주장값과 비교하는 방법입니다.
We ran 5 replicates on 5 days; does our analyser meet the manufacturer's precision claims?
5일 동안 하루 5번씩 측정했는데, 우리 분석기가 제조사가 밝힌 정밀도를 만족하나요?
A one-way random-effects analysis of variance with runs as groups gives the repeatability variance and the between-run variance, which add up to the within-laboratory variance with Satterthwaite degrees of freedom. Each estimate is compared with the claimed value through an upper verification limit based on the chi-square distribution, and the mean can be compared with a reference value to estimate bias.
When to use Single-Laboratory Precision Verification (CLSI EP15)
- A laboratory verifies the precision claims of a method or analyser before routine use.
- Replicates were measured over several days or runs at one concentration level.
- You also want a bias estimate against a reference material.
When to use something else
- Several laboratories took part and you want reproducibility; use an interlaboratory precision study (ISO 5725-2).
- You are establishing precision from scratch with a long design; use a full variance components analysis (EP05-type).
Assumptions and requirements
- Run effects and within-run errors are independent and approximately normal.
- The within-run variance is the same in every run.
- The design is balanced for exact bias intervals (unbalanced designs use an approximation).
What the analysis reports
- Repeatability, between-run and within-laboratory SD and CV with chi-square CIs
- Satterthwaite degrees of freedom for the within-laboratory SD
- Upper verification limits and pass/fail against claimed SDs or CVs
- Bias against a reference value with CI and verification interval
How to report it
Across [value] runs (N = [value]), the repeatability SD was [value] (CV = [value]%) and the within-laboratory SD was [value] (CV = [value]%, df = [value]), both below the upper verification limits of [value] and [value].
Related methods
- Variance Components Analysis (Random-Effects ANOVA)
- Interlaboratory Precision Study (ISO 5725-2)
- Intraclass Correlation (ICC)
How this method is validated
이 방법의 검증 근거
- Automated tests. Checked by automated run tests; no external reference comparison is recorded yet.
자동 시험. 자동 실행 시험을 거칩니다. 외부 참조값과 대조한 기록은 아직 없습니다.
Run Single-Laboratory Precision Verification (CLSI EP15) on Quorant
- Data you need
- a table (CSV or Excel) with an outcome variable and a grouping variable
- Price
- 400 credits for one analysis up to 10,000 rows (complexity level 3); larger or wider data costs more. Optional methodologist review: 1800 credits.
- Checking
- Automated run tests; no external reference comparison yet
- Before you pay
- A free pre-analysis diagnosis checks your data and gives an exact quote; nothing is deducted until you approve it.